GDS-50秒表檢定儀是一用的時間檢定儀,供各級計量部門,工廠,院校及科研單位檢定各種秒表、毫秒儀,機械秒表等,性能符合檢定規(guī)程的要求。
技術指標:
1、 檢定電子秒表和機械秒表:
在1-99999秒內(nèi)可任意置數(shù)
準確度±(2*10-7*輸出時段)秒±0.003秒
2、 檢定電子毫秒儀(正負兩組輸出)
在1-99999秒內(nèi)可任意置數(shù),小輸出100微秒
準確度±(2*10-7*輸出時段)秒±0.003秒
3、 檢定電秒表(適合新規(guī)程JJG237-95)
在1-99999秒內(nèi)可 任意置數(shù)
指針式準確度:電同步<±0.6毫秒
數(shù)字式準確度:±(2*10-7*輸出時段)秒±0.006秒
Introduction
The calibrator is a precision instrument calibrating mechanical stopwatch, electronicstopwatch, and 405/407/408/415/417 type pointer electric stopwatch. The SF2001stopwatch calibrator is compliant with the stopwatch calibration procedure JJG 237-2010 issued by the National Metrological Bureau. It could be used at metrologicaldepartment at all levels, factory, institutions and all research institutes to calibrateelectric stopwatch (405,407,408,415/417), mechanical stopwatch and digital quartzelectronic stopwatch. The instrument is fully digitally controlled, high reliability,high accuracy, complete functions, convenient operation, and strong anti-interfere
ability.
GDS-50 stopwatch calibrator and its accessories:
1)GDS-50 stopwatch calibrator 1 piece
2)Three-core power cord 1 piece
3)Dual banana plug test lead 2 pieces
4)BNC-banana test lead 1 piece
5)Soft copy of User’s Guide 1 piece
Main Parameter
1.1 Crystal Oscillator
Nominal frequency: 10MHz
Daily aging rate: ≤5×10 -9 /day
Second stability: ≤5×10 -10 /s
Accuracy: ≤1×10 -7
Warming-up time: over 2 hours
1.2 Calibrate mechanical stopwatch and electronic stopwatch(T0 is inputtedcalibration period of time)
Input range: T 0 :1s-99999s
Accuracy: better than ±(1×10 -7 ×T 0 +3ms)
1.3 Calibrate electric stopwatch
Pointer:
Input range:
Continuous: T 0 :0.1s-99999s
Trigger state: T 0 :1s-99999s
Pause state: T 0 :0.1s-99999s
Accuracy: better than ±(commercial power frequency accuracy×T 0 +0.6ms)
1.4 Calibrate millisecond counter and digital electric stopwatch
Input range: T 0 :1×10 2 μs-1×10 9 μs
Accuracy: better than ±(1×10 -7 ×T 0 +3μs)
The SF2001 calibrator is fully compliant with the mechanical stopwatch and digital quartzstopwatch calibration procedure, electric stopwatch calibration procedure, and electronicmillisecond counter calibration procedure, issued by the National Metrological Bureau.
2.1 Working conditions
Voltage: 220(1±10%)V;
Frequency: 50(1±5%)Hz;
Power consumption: <60VA;
Temperature : 0~40℃;
Humidity: <80%
2.2 Mechanical characteristics
Dimension: 329mm×283mm×155mm Weight: 3.7kg
2.3 Operation characteristics: Input with keys, menu display
2.4 Display mode: LCD display,English menu
2.5 Technique: With SMT technique and large-scale IC, the calibrator is of highreliability, small volume and long service life.